TECHNICAL DETAILS - FALCON RANGE 

MEASUREMENT UNCERTAINTY 
 
(X,Y) 150mm x 100mm measuring stage 
 
Uncertainty formula U952D = 7+(6.5L/1000)µm, where L = length in mm, using controlled conditions. 
 
(X,Y) 150mm x 150mm measuring stage 
 
Uncertainty formula U952D = 4+(5.5L/1000)µm, where L = length in mm, using controlled conditions. 
 
(X,Y) 150mm x 150mm CNC measuring stage 
 
Uncertainty formula U952D = 4+(5.5L/1000)µm, where L = length in mm, using controlled conditions. 
 
(Z) 
 
Z-axis accuracy 10µm, using highest available magnification under controlled conditions. 
VIDEO CAMERA 
 
1.3MP colour CMOS camera with 5-position indexed zoom (5:1 ratio) and iris aperture control. 
 
ILLUMINATION 
 
Quadrant controllable surface illumination provided by 20 x high intensity LEDs 
Controllable substage illumination provided by 1 x high intensity LED 
MEASURING RANGE (X,Y,Z) 
 
150mm x 100mm x 125mm (10kg max load) 
150mm x 150mm x 115mm (15Kg max load) 
Falcon systems at a glance 
*QC-300 (Touch-screen video microprocessor) * QC-5000 (PC-based messurement software) 